Static and dynamic characteristics, first- and second-order response, transducers, thermocouples and RTDs, pressure and vacuum gauges, chromatography and spectroscopy, level, BET and electron microscopy, ISA symbols and P&IDs from the ASTU CHE241405 syllabus.
What is inside
- First- and second-order instrument response
- Thermocouples, RTDs, pyrometers and pressure gauges
- Chromatography, spectroscopy and level measurement
- BET surface area, SEM/TEM and P&ID symbols
Syllabus: ASTU CHE241405 Process Instrumentation, semester 4, NEP 2020 (2024-25 onwards). Official syllabus PDF.
Elements, static and dynamic characteristics of instruments
Topics: Elements of instruments; static characteristics; dynamic characteristics; linear ODEs and the Laplace transform in instrumentation; response of first- and second-order instruments.
Elements of an instrument and its static characteristics
Every measuring instrument can be broken into functional elements: a primary sensing element that responds to the measured variable (a thermocouple junction, a Bourdon tube, an orifice), a variable-conversion element that changes the signal into a more convenient form (strain to resistance, displacement to voltage), a variable-manipulation element that amplifies or scales it, a data-transmission element (pneumatic tubing at 0.2–1 bar, electrical 4–20 mA, digital fieldbus), and a data-presentation element (indicator, recorder, display, controller input).
Static characteristics describe the instrument once the reading has settled. Accuracy is closeness to the true value; precision (repeatability) is closeness of repeated readings to each other. Sensitivity is output change per unit input; resolution the smallest detectable change; threshold the smallest input that gives any output; dead zone the range over which the input can change without output change. Linearity is the closeness of the calibration curve to a straight line; hysteresis the difference between readings taken with increasing and decreasing input; drift a slow change of reading with time. Errors are gross (blunders), systematic (instrumental, environmental, observational; removable by calibration) or random (treated statistically). Static characteristics are found by calibration against a standard of higher accuracy.
- Accuracy ≠ precision.
- Sensitivity = Δoutput/Δinput.
- Hysteresis: up and down readings differ.
- Errors: gross, systematic, random.
Dynamic characteristics: first- and second-order instruments
When the measured variable changes, an instrument does not respond instantly. Its behaviour is described by a linear differential equation, solved most easily with the Laplace transform, which turns derivatives into multiplication by s and gives a transfer function G(s) = output/input. A zero-order instrument (a potentiometer) follows the input exactly. A first-order instrument obeys τ dy/dt + y = Kx, G(s) = K/(τs + 1). A mercury thermometer is the classic case: an energy balance mC dT_b/dt = hA(T − T_b) gives τ = mC/hA.
For a step of size M, y = KM(1 − e^(−t/τ)): 63.2 % of the change after τ, 86.5 % after 2τ, 95 % after 3τ, 98.2 % after 4τ. For a ramp x = rt, after the transient the reading lags by a constant time τ (error rτ). For a sinusoid of frequency ω, the reading has amplitude ratio 1/√(1 + ω²τ²) and lags by tan⁻¹(ωτ).
A second-order instrument obeys τ²ÿ + 2ζτẏ + y = Kx, G(s) = K/(τ²s² + 2ζτs + 1), with natural frequency ω_n = 1/τ and damping ratio ζ: spring–mass systems, U-tube manometers, a thermometer in a thermowell. For ζ < 1 the step response overshoots and oscillates: overshoot = exp(−πζ/√(1 − ζ²)), decay ratio = overshoot², period 2π/(ω_n√(1 − ζ²)), settling time ≈ 4/(ζω_n). ζ = 1 is critically damped (fastest without overshoot); ζ > 1 overdamped. Instruments are usually designed with ζ ≈ 0.6–0.7.
- First order: G = K/(τs + 1); 63.2 % at τ.
- Ramp: constant lag τ; sine: AR = 1/√(1 + ω²τ²).
- Second order: G = K/(τ²s² + 2ζτs + 1).
- Overshoot = e^(−πζ/√(1−ζ²)); decay ratio = OS².
Example: A thermometer with τ = 10 s at 25 °C is plunged into a bath at 100 °C. What does it read after 15 s, and when does it reach 99 % of the change?
- y = 25 + 75(1 − e^(−15/10)) = 25 + 75 × 0.7769 = 83.3 °C.
- 99 %: t = −τ ln(0.01) = 4.605 × 10 = 46 s.
Answer: 83.3 °C after 15 s; 46 s to 99 %.
Example: A second-order instrument has ζ = 0.4 and ω_n = 5 rad/s. Find the overshoot, decay ratio, period and settling time.
- √(1 − ζ²) = 0.9165.
- Overshoot = exp(−π × 0.4/0.9165) = exp(−1.371) = 0.254 (25.4 %); decay ratio = 0.064.
- Period = 2π/(5 × 0.9165) = 1.37 s; settling ≈ 4/(0.4 × 5) = 2 s.
Answer: 25.4 % overshoot, decay ratio 0.064, period 1.37 s, settles in about 2 s.
Example: Derive the transfer function of a mercury thermometer.
- Energy balance on the bulb: mC dT_b/dt = hA(T − T_b).
- Define τ = mC/hA: τ dT_b/dt + T_b = T.
- Laplace transform with deviation variables (zero initial values): τs T_b(s) + T_b(s) = T(s).
- G(s) = T_b(s)/T(s) = 1/(τs + 1).
Answer: A first-order lag with τ = mC/hA.
Process dynamics and control
A first-order system responds to a step by reaching 63.2 % of the final change after one time constant τ, 95 % after 3τ and 98 % after 4τ. Controllers: on–off (simple, cycles), proportional (leaves an offset), PI (integral removes offset), PID (derivative adds anticipation). Proportional band PB = 100/Kc %.
Feedback control corrects after an error appears; feed-forward measures a disturbance and acts before it affects the output. Control valves are air-to-open (fail closed) or air-to-close (fail open), chosen for safety.
- 63.2 % at τ; 95 % at 3τ
- P leaves offset; I removes it
- PB = 100/Kc %
Example: A thermometer with τ = 5 s is plunged into hot water. What fraction of the change does it show after 15 s?
- 15 s = 3τ.
- 1 − e⁻³ ≈ 0.95 = 95 %.
Answer: 95 %
Example: A proportional controller has a proportional band of 50 %. Find its gain.
- Kc = 100/PB.
- = 100/50 = 2.
Answer: 2
Functional elements: primary sensing, variable conversion, manipulation, transmission, presentation. Static characteristics: accuracy, precision, sensitivity, resolution, threshold, dead zone, linearity, hysteresis, drift; errors gross, systematic, random; calibration. Dynamic: zero-, first- and second-order models via Laplace transforms. First order: τ, 63.2 % at τ, ramp lag τ, sinusoidal AR and phase lag. Second order: ω_n and ζ; overshoot, decay ratio, period, settling time.
- Thermometer τ = mC/hA: smaller bulb, faster response.
- A ramp is followed with a constant lag of τ.
- ζ ≈ 0.6–0.7 is the usual design target.
- Laplace: d/dt → s (zero initial deviation).
Key formulas
- τ dy/dt + y = Kx ; G(s) = K/(τs + 1)
- y = KM(1 − e^(−t/τ))
- AR = 1/√(1 + ω²τ²), φ = −tan⁻¹(ωτ)
- OS = exp(−πζ/√(1 − ζ²))
- G(s) = K/(τ²s² + 2ζτs + 1)
Practice questions
A first-order instrument reaches what fraction of a step change in one time constant?
- 50 %
- 63.2 %
- 86.5 %
- 95 %
Show the answer
Answer: B. 63.2 % 1 − e⁻¹.
The time constant of a thermometer bulb is
- hA/mC
- mC/hA
- m/hA
- hA/m
Show the answer
Answer: B. mC/hA Energy balance.
A second-order system with ζ < 1 is
- overdamped
- critically damped
- underdamped
- unstable
Show the answer
Answer: C. underdamped Oscillates.
Hysteresis is
- random scatter
- difference between increasing and decreasing readings
- slow change with time
- smallest detectable change
Show the answer
Answer: B. difference between increasing and decreasing readings Static characteristic.
The decay ratio of an underdamped second-order response equals
- overshoot
- overshoot squared
- 1/overshoot
- ζ²
Show the answer
Answer: B. overshoot squared Successive peaks.
Theory questions
Derive the response of a first-order instrument to a step and to a ramp input. (6 marks)
Show the model answer
Model τ dy/dt + y = Kx, G(s) = K/(τs + 1). Step x = M: X(s) = M/s, Y(s) = KM/(s(τs + 1)) = KM[1/s − τ/(τs + 1)] ⇒ y = KM(1 − e^(−t/τ)); 63.2 % at t = τ. Ramp x = rt: X(s) = r/s², Y(s) = Kr/(s²(τs + 1)) = Kr[1/s² − τ/s + τ²/(τs + 1)] ⇒ y = Kr(t − τ + τe^(−t/τ)). For K = 1, after a few τ, y ≈ r(t − τ): the output follows the input with a constant time lag τ and a dynamic error rτ.
Define the static characteristics of an instrument. (5 marks)
Show the model answer
Accuracy: closeness of the reading to the true value, as ±% of full scale or of reading. Precision/repeatability: closeness of repeated readings of the same input. Sensitivity: ratio of output change to input change. Resolution: smallest input change that produces a detectable output change. Threshold: smallest input from zero that gives a detectable output. Dead zone: largest input change with no output. Linearity: maximum deviation of the calibration curve from a best straight line. Hysteresis: maximum difference between readings for increasing and decreasing input. Drift: change in reading with time at constant input (zero drift, span drift). Range and span: limits and their difference.
Transducer elements
Topics: Transducer elements; different types of transducers.
What a transducer does and the main types
A transducer converts one form of energy or signal into another, usually into an electrical signal that can be amplified, transmitted and recorded. Active (self-generating) transducers produce their own output from the measured energy: thermocouples (Seebeck emf), piezoelectric crystals (charge from force), photovoltaic cells, tachogenerators. Passive transducers change an electrical property and need an external supply: resistive (strain gauges, RTDs, thermistors, potentiometers), capacitive and inductive (LVDT, variable reluctance) types. Transducers may also be classed as analogue or digital, primary or secondary, and as transducers or inverse transducers (electrical to mechanical, such as actuators).
Resistive: a strain gauge’s resistance changes with strain as ΔR/R = GF × ε (gauge factor GF ≈ 2 for metal foil, up to 100+ for semiconductors), read in a Wheatstone bridge. A potentiometer gives a voltage proportional to wiper position. Capacitive: C = ε₀ε_rA/d, so changing plate spacing, area or dielectric (level, humidity, pressure diaphragms) changes C. Inductive: the linear variable differential transformer (LVDT) has a primary coil and two secondaries in opposition; a movable core gives an output voltage proportional to displacement with a phase showing direction. Piezoelectric: quartz or PZT produce charge q = dF (d ≈ 2.3 pC/N for quartz), ideal for dynamic forces and vibration but not static loads. Others: Hall-effect sensors, photoconductive cells, ultrasonic and optical-fibre transducers.
Transmitters package a sensor with electronics and send a standard signal: 4–20 mA (live zero so a broken wire reads 0 mA, not 0 %), 3–15 psi (0.2–1 bar) pneumatic, or digital (HART, Foundation Fieldbus, Profibus, wireless).
- Active: thermocouple, piezoelectric, photovoltaic.
- Passive: resistive, capacitive, inductive (need excitation).
- Strain gauge: ΔR/R = GF·ε.
- LVDT: voltage ∝ core displacement.
- 4–20 mA live zero.
Example: A 120 Ω strain gauge (GF = 2) feels a strain of 500 µε. Find ΔR and the output of a quarter bridge excited at 10 V.
- ΔR/R = 2 × 500 × 10⁻⁶ = 0.001 ⇒ ΔR = 0.12 Ω.
- Quarter bridge: V_o ≈ E × GF × ε/4 = 10 × 0.001/4.
Answer: ΔR = 0.12 Ω; output ≈ 2.5 mV, which needs amplification.
Example: A quartz crystal (d = 2.3 pC/N) with total capacitance 1000 pF is loaded with 100 N. Find the charge and the open-circuit voltage.
- q = dF = 2.3 × 100 = 230 pC.
- V = q/C = 230 × 10⁻¹²/1000 × 10⁻¹².
Answer: 230 pC and 0.23 V (which leaks away with time, so static forces cannot be measured).
Example: A transmitter ranged 0–10 bar gives 4–20 mA. What pressure does 13.6 mA mean?
- Fraction = (13.6 − 4)/(20 − 4) = 9.6/16 = 0.6.
- p = 0.6 × 10 = 6 bar.
Answer: 6 bar.
Transducers convert a measured quantity into another (usually electrical) signal. Active: thermocouple, piezoelectric, photovoltaic, tachogenerator. Passive: resistive (strain gauge ΔR/R = GF·ε, potentiometer, RTD, thermistor), capacitive (C = ε₀ε_rA/d), inductive (LVDT, reluctance). Also Hall-effect, photoconductive, ultrasonic and fibre-optic types. Transmitters send 4–20 mA, 0.2–1 bar pneumatic or digital (HART, fieldbus) signals.
- Wheatstone bridge reads small resistance changes.
- Piezoelectric: dynamic only.
- LVDT: frictionless, infinite resolution.
- Live zero detects broken wires.
Key formulas
- ΔR/R = GF × ε
- C = ε₀ε_r A/d
- q = dF
- p = range × (I − 4)/16 (4–20 mA)
Practice questions
A thermocouple is a
- passive transducer
- active transducer
- capacitive transducer
- inverse transducer
Show the answer
Answer: B. active transducer Self-generating emf.
The gauge factor of a metal foil strain gauge is about
- 0.2
- 2
- 20
- 200
Show the answer
Answer: B. 2 Metals ≈ 2.
An LVDT measures
- temperature
- displacement
- flow
- pH
Show the answer
Answer: B. displacement Core position.
Piezoelectric transducers are unsuitable for
- vibration
- impact
- static force
- acoustic pressure
Show the answer
Answer: C. static force Charge leaks.
A 4–20 mA transmitter reads 12 mA. The measurement is at
- 25 %
- 50 %
- 60 %
- 75 %
Show the answer
Answer: B. 50 % (12 − 4)/16.
Theory questions
Classify transducers with examples. (5 marks)
Show the model answer
Active (self-generating) vs passive (externally powered): thermocouple, piezoelectric, photovoltaic, tachogenerator vs strain gauge, RTD, potentiometer, capacitive, LVDT. Primary vs secondary: Bourdon tube (pressure → displacement) followed by LVDT (displacement → voltage). Analogue vs digital: strain gauge vs shaft encoder. Transducer vs inverse transducer: microphone vs loudspeaker/actuator. By principle: resistive, capacitive, inductive, piezoelectric, thermoelectric, photoelectric, Hall effect, ultrasonic.
Explain the working of an LVDT with a sketch. (5 marks)
Show the model answer
A primary coil excited by AC (1–10 kHz) sits between two identical secondary coils connected in series opposition, all on a hollow former with a movable soft-iron core attached to the measured member. With the core central the induced voltages in the secondaries are equal and cancel (null). Moving the core couples more flux to one secondary, so the output voltage rises linearly with displacement; its phase relative to the primary shows the direction. A phase-sensitive demodulator gives a DC signal. Advantages: no friction or contact, high resolution, rugged, linear over the range; disadvantages: AC excitation, sensitive to stray magnetic fields.
Temperature measurement
Topics: Thermocouples, resistance thermometers, radiation pyrometers, optical pyrometers, photoelectric pyrometers, thermistors.
Thermocouples, RTDs and thermistors
A thermocouple is two dissimilar metal wires joined at a hot (measuring) junction; the circuit’s emf depends on the temperature difference between the hot and reference (cold) junctions (Seebeck effect). Standard types: K (chromel–alumel, −200 to 1260 °C, about 41 µV/K, general purpose), J (iron–constantan, to 760 °C), T (copper–constantan, cryogenic to 370 °C), E (chromel–constantan, highest output), R and S (platinum–rhodium, to about 1600 °C), B (to 1800 °C). Laws of thermocouples: of homogeneous circuits, of intermediate metals (a copper meter in the circuit adds no emf if its junctions are at equal temperature) and of intermediate temperatures (E(T, 0) = E(T, T_cj) + E(T_cj, 0), the basis of cold-junction compensation). Compensating or extension cables carry the signal to the instrument; thermowells protect the junction but slow the response.
Resistance temperature detectors use the rise in resistance of a pure metal, usually platinum: R = R₀(1 + At + Bt²) (Callendar–Van Dusen), about 0.385 Ω/K for a Pt100. They are more accurate and stable than thermocouples but slower and limited to about 850 °C. They are read in a bridge; 3- and 4-wire connections cancel lead-wire resistance; self-heating from the measuring current must be kept small. Thermistors are semiconductor (metal-oxide) resistors whose resistance falls steeply with temperature (NTC): R = R₀ exp[β(1/T − 1/T₀)], several % per kelvin, very sensitive over narrow ranges (−50 to 150 °C), non-linear, cheap and small.
- Type K ≈ 41 µV/K; J, T, E, R, S, B other ranges.
- Law of intermediate temperatures → cold-junction compensation.
- Pt100: 100 Ω at 0 °C, 138.5 Ω at 100 °C.
- Thermistor: R = R₀ e^(β(1/T − 1/T₀)).
Example: A type K thermocouple with its cold junction at 25 °C gives 20.0 mV. Find the hot-junction temperature (E(25 °C) = 1.000 mV).
- E(T, 0) = 20.0 + 1.000 = 21.0 mV.
- From the type K table: 20.644 mV at 500 °C, sensitivity ≈ 0.0427 mV/K there.
- T ≈ 500 + (21.0 − 20.644)/0.0427 ≈ 508.3 °C.
Answer: About 508 °C (ignoring the cold junction would give about 485 °C).
Example: A Pt100 reads 138.5 Ω in a 2-wire circuit with 0.5 Ω total lead resistance. Find the temperature.
- Sensor R = 138.5 − 0.5 = 138.0 Ω.
- Solve 100(1 + 3.9083 × 10⁻³ t − 5.775 × 10⁻⁷ t²) = 138.0.
- t ≈ 98.7 °C.
Answer: 98.7 °C (the leads alone cause a 1.3 °C error).
Radiation, optical and photoelectric pyrometers
Above about 600 °C, or for moving or inaccessible objects, temperature is measured from emitted thermal radiation without contact. A total-radiation pyrometer focuses radiation from the target with a lens or mirror onto a thermopile; by the Stefan–Boltzmann law the energy is εσT⁴, so the instrument reads correctly only for the emissivity it is set for. A body of emissivity ε read by a pyrometer calibrated for a black body gives an apparent temperature T_a with T = T_a/ε^(1/4).
The disappearing-filament optical pyrometer compares the brightness of the target at one wavelength (red, about 0.65 µm) with an electrically heated tungsten filament seen against it; the filament current at which it disappears against the image gives the temperature, typically 700–3000 °C. Photoelectric (infrared) pyrometers replace the eye with a photodetector (silicon, InGaAs, PbS), are fast and automatic, and are now the usual non-contact thermometers; two-colour (ratio) pyrometers measure at two wavelengths and largely cancel emissivity and dust effects.
- Total radiation: E = εσT⁴; T = T_a/ε^(1/4).
- Optical pyrometer: brightness match at 0.65 µm.
- Photoelectric/IR: fast, automatic.
- Ratio pyrometer: emissivity largely cancels.
Example: A total-radiation pyrometer calibrated for a black body reads 1000 K on a surface of emissivity 0.8. What is the true temperature?
- εσT⁴ = σT_a⁴ ⇒ T = T_a/ε^(1/4).
- ε^(1/4) = 0.8^0.25 = 0.9457.
- T = 1000/0.9457 = 1057 K.
Answer: 1057 K: the pyrometer under-reads by 57 K.
Measuring temperature and pressure
Thermocouples produce a voltage from two dissimilar metals (Seebeck effect); types K and J are common. Resistance thermometers (RTDs, e.g. Pt100) follow R = R₀(1 + αT) with α ≈ 0.00385/°C for platinum. Thermistors are very sensitive but non-linear. Pyrometers measure hot bodies by their radiation.
Pressure: manometers for low pressures, Bourdon tubes for gauges, diaphragms and strain-gauge transmitters. Absolute pressure = gauge pressure + atmospheric pressure.
- RTD: R = R₀(1 + αT)
- Thermocouple: Seebeck voltage
- p_abs = p_gauge + p_atm
Example: A Pt100 RTD (α = 0.00385/°C) is at 100 °C. Find its resistance.
- R = 100(1 + 0.00385 × 100).
- = 138.5 Ω.
Answer: 138.5 Ω
Example: A gauge reads 2 bar. Find the absolute pressure.
- p_abs = 2 + 1.013.
- ≈ 3.01 bar.
Answer: ≈ 3.01 bar (abs)
Thermocouples (Seebeck emf; types K, J, T, E, R, S, B; laws of homogeneous circuits, intermediate metals and intermediate temperatures; cold-junction compensation). RTDs (platinum, Callendar–Van Dusen, bridges, 3/4-wire). Thermistors (NTC, β equation, sensitive, non-linear). Non-contact: total-radiation pyrometers (Stefan–Boltzmann, emissivity correction), disappearing-filament optical pyrometers, photoelectric/IR and two-colour pyrometers.
- Thermowells protect but slow the response.
- RTD: accurate, stable, up to ~850 °C.
- Pyrometers read low on low-emissivity surfaces.
- Two-colour pyrometers tolerate smoke and dust.
Key formulas
- E(T, 0) = E(T, T_cj) + E(T_cj, 0)
- R = R₀(1 + At + Bt²)
- R = R₀ exp[β(1/T − 1/T₀)]
- T = T_a/ε^(1/4)
Practice questions
Which thermocouple is best for cryogenic temperatures?
- Type T
- Type B
- Type S
- Type R
Show the answer
Answer: A. Type T Copper–constantan.
Cold-junction compensation is based on the law of
- homogeneous circuits
- intermediate metals
- intermediate temperatures
- Ohm
Show the answer
Answer: C. intermediate temperatures Adding emfs.
A Pt100 has a resistance of 100 Ω at
- 0 °C
- 25 °C
- 100 °C
- −273 °C
Show the answer
Answer: A. 0 °C Definition.
An NTC thermistor’s resistance
- rises with temperature
- falls with temperature
- is constant
- is zero above 100 °C
Show the answer
Answer: B. falls with temperature Semiconductor.
An optical pyrometer works by comparing
- total radiation
- brightness at one wavelength
- thermal expansion
- emf
Show the answer
Answer: B. brightness at one wavelength Disappearing filament.
Theory questions
State the laws of thermocouples and explain cold-junction compensation. (5 marks)
Show the model answer
Law of homogeneous circuits: no emf can be produced in a circuit of a single homogeneous metal by heating alone. Law of intermediate metals: inserting a third metal does not change the emf if both its junctions are at the same temperature (so meters and copper leads can be used). Law of intermediate temperatures: E(T₁, T₃) = E(T₁, T₂) + E(T₂, T₃). Tables are for a 0 °C reference. If the reference is at ambient T_cj, the measured E(T, T_cj) is corrected by adding E(T_cj, 0), either from a measurement of T_cj (thermistor or RTD at the terminals) or by a bridge circuit that generates the equivalent emf automatically.
Compare thermocouples, RTDs and thermistors. (5 marks)
Show the model answer
Thermocouple: self-powered, wide range (−200 to 1800 °C), rugged, fast, cheap; low output (µV/K), needs cold-junction compensation, less accurate (±1–2 °C). RTD (Pt): accurate (±0.1–0.3 °C), stable, nearly linear, −200 to 850 °C; needs excitation, slower, costlier, lead-wire and self-heating errors. Thermistor: very high sensitivity (several %/K), small and fast, cheap; narrow range (−50 to 150 °C), highly non-linear, less interchangeable and stable.
Measurement of pressure and vacuum
Topics: Bourdon tube, diaphragm pressure gauge, bellows gauge, pressure spring; McLeod gauge, Pirani gauge, ionisation gauge, thermocouple gauge.
Elastic pressure elements
Most industrial pressure gauges use an elastic element that deflects under pressure. The Bourdon tube is a flattened (oval) tube bent into a C, spiral or helix with one end fixed to the pressure inlet and the other closed and free. Pressure inside tends to make the cross-section circular, which straightens the tube; the tip moves almost in proportion to pressure and drives a pointer through a sector and pinion. C-tubes cover about 0.5 to several thousand bar; spirals and helices give more movement for recorders. Materials: phosphor bronze, stainless steel, beryllium copper.
Diaphragm gauges use a thin flat or corrugated metal (or non-metal) disc clamped at its edge; the centre deflects with pressure (for a flat plate y = 3(1 − ν²)pR⁴/(16Et³) in the linear range). They suit low pressures, corrosive or slurry service and, with a strain gauge or capacitance pickup, form the sensing element of most electronic pressure transmitters. Bellows are thin-walled corrugated cylinders that expand axially, giving large movement at low pressure, often opposed by a spring (the “pressure spring”). Manometers and dead-weight testers serve as standards for calibrating these gauges.
- Bourdon: oval tube straightens under pressure.
- Diaphragm: low pressure, corrosive service, transmitters.
- Bellows: large stroke at low pressure.
- Dead-weight tester: p = mg/A, calibration standard.
Example: A dead-weight tester has a 10 mm piston loaded with 20 kg. What pressure does it generate?
- A = π × 0.01²/4 = 7.854 × 10⁻⁵ m².
- p = mg/A = 20 × 9.80665/7.854 × 10⁻⁵ = 2.497 × 10⁶ Pa.
Answer: About 24.97 bar.
Vacuum gauges: McLeod, Pirani, thermocouple and ionisation
Below about 1 mmHg elastic elements barely move, so other principles are used. The McLeod gauge traps a known volume V of the gas at the unknown pressure and compresses it into a fine closed capillary; by Boyle’s law p = ah²/(V − ah) ≈ ah²/V, where a is the capillary area and h the mercury difference. It is an absolute standard (about 10⁻⁵–1 mmHg) but gives only intermittent readings and condensable vapours spoil it.
Thermal-conductivity gauges use the fact that, at low pressure, the heat a gas conducts away from a hot wire falls with pressure. In the Pirani gauge the wire is one arm of a Wheatstone bridge and its resistance (temperature) indicates pressure; in the thermocouple gauge a thermocouple welded to the heated wire measures its temperature directly. Both cover about 10⁻³–1 mmHg and must be calibrated for the gas. Ionisation gauges measure still lower pressures: in the hot-cathode type, electrons from a filament ionise gas molecules and the ion current collected is proportional to pressure (about 10⁻³ down to 10⁻¹⁰ mmHg); the cold-cathode (Penning) type uses a high voltage and magnetic field instead of a filament.
- McLeod: p ≈ ah²/V (absolute standard).
- Pirani and thermocouple gauges: heat loss from a hot wire.
- Ionisation: ion current ∝ pressure, down to 10⁻¹⁰ mmHg.
- Thermal and ionisation gauges depend on the gas.
Example: A McLeod gauge has V = 100 cm³ and a 0.5 mm capillary. The reading is h = 2 cm. Find the pressure.
- a = π(0.05 cm)²/4 = 1.963 × 10⁻³ cm².
- p = ah²/(V − ah) = 1.963 × 10⁻³ × 4/(100 − 0.0039) cm Hg.
- = 7.85 × 10⁻⁵ cm Hg = 7.85 × 10⁻⁴ mmHg.
Answer: About 7.9 × 10⁻⁴ mmHg (0.105 Pa).
Elastic elements: Bourdon tube (C, spiral, helical; oval section straightens), diaphragms (flat or corrugated; transmitters), bellows and pressure springs; calibrated with manometers and dead-weight testers. Vacuum: McLeod gauge (Boyle compression, absolute standard), Pirani and thermocouple gauges (hot-wire heat loss), hot- and cold-cathode ionisation gauges for high vacuum.
- Gauge pressure = absolute − atmospheric.
- Use a siphon or snubber to protect Bourdon gauges from steam and pulsation.
- Thermal gauges need gas-specific calibration.
- Ionisation gauges work only at high vacuum.
Key formulas
- p = mg/A (dead weight)
- y = 3(1 − ν²)pR⁴/(16Et³)
- p = ah²/(V − ah) (McLeod)
Practice questions
A Bourdon tube has a cross-section that is
- circular
- oval
- square
- triangular
Show the answer
Answer: B. oval Straightens under pressure.
The McLeod gauge is based on
- Boyle’s law
- Charles’ law
- Ohm’s law
- Bernoulli
Show the answer
Answer: A. Boyle’s law Compression of trapped gas.
A Pirani gauge measures pressure from
- ion current
- heat loss from a hot wire
- deflection of a tube
- mercury height
Show the answer
Answer: B. heat loss from a hot wire Thermal conductivity.
Which gauge measures the lowest pressures?
- Bourdon
- McLeod
- Pirani
- Ionisation
Show the answer
Answer: D. Ionisation Down to 10⁻¹⁰ mmHg.
A dead-weight tester is used to
- measure vacuum
- calibrate pressure gauges
- measure flow
- measure level
Show the answer
Answer: B. calibrate pressure gauges Primary standard.
Theory questions
Explain the construction and working of a Bourdon tube pressure gauge. (5 marks)
Show the model answer
A tube of oval/elliptical cross-section is bent into a C (≈ 270°). The open end is fixed to the socket connected to the pressure source; the other end is sealed and free. When pressure acts inside, the oval section tends to become circular, which increases the radius of curvature, so the free end moves outward almost linearly with pressure. The tip movement is magnified by a link, toothed sector and pinion that turns the pointer over a calibrated dial; a hairspring removes backlash. Ranges from a fraction of a bar to several thousand bar; spiral and helical forms give larger tip travel. Over-range, vibration and temperature cause errors; siphons and snubbers protect it.
Describe the McLeod and Pirani vacuum gauges and compare them. (6 marks)
Show the model answer
McLeod: gas at unknown pressure p fills a bulb of volume V; raising the mercury reservoir traps it and compresses it into a sealed capillary of area a; the mercury in the reference capillary rises h above that in the closed one, so p V = (p + h)ah, giving p = ah²/(V − ah). Absolute, used as a standard, range ≈ 10⁻⁵–1 mmHg; manual, intermittent, unsuitable for condensable vapours. Pirani: a heated wire (one arm of a Wheatstone bridge) loses heat by gas conduction proportional to pressure at low pressure; as pressure falls the wire gets hotter and its resistance rises, unbalancing the bridge (or a constant-temperature circuit measures power). Range ≈ 10⁻³–1 mmHg; continuous, electrical output, but must be calibrated for each gas (against a McLeod).
Composition measuring instruments
Topics: Gas chromatography, liquid chromatography, UV–Vis spectroscopy, mass spectrometry, ion meters, fuel gas analysis, oxygen analysis, polarography, colorimetry, combustible gas analysis.
Chromatography and mass spectrometry
Chromatography separates a mixture by passing it in a mobile phase over a stationary phase; components that interact more with the stationary phase move more slowly. In gas chromatography (GC) a carrier gas (He, N₂, H₂) sweeps a vapourised sample through a packed or capillary column in a temperature-controlled oven; detectors include the thermal conductivity detector (TCD, universal), flame ionisation detector (FID, very sensitive to hydrocarbons) and electron capture detector (ECD, halogens). In liquid chromatography (HPLC) a liquid mobile phase is pumped at high pressure through a column of fine particles (often reversed-phase C18); UV, refractive-index and fluorescence detectors are common. Process GCs analyse plant streams automatically every few minutes.
Each peak’s retention time identifies a component (compared with standards) and its area gives the amount, after correcting with response factors. Column performance is described by the plate number N = 16(t_R/w)² and plate height HETP = L/N; separation of two peaks by the resolution R_s = 2(t₂ − t₁)/(w₁ + w₂), with R_s ≥ 1.5 for baseline separation. A mass spectrometer ionises molecules (electron impact), separates the ions by mass-to-charge ratio in a quadrupole, magnetic-sector or time-of-flight analyser, and counts them; the fragmentation pattern identifies compounds. Coupled GC–MS is the standard for identifying unknowns, and process MS analysers measure many gases in seconds.
- GC detectors: TCD (universal), FID (hydrocarbons), ECD (halogens).
- N = 16(t_R/w)²; HETP = L/N.
- R_s = 2(t₂ − t₁)/(w₁ + w₂) ≥ 1.5.
- MS separates ions by m/z.
Example: Two GC peaks elute at 4.2 and 4.8 min with base widths 0.30 and 0.32 min on a 30 m column. Find the resolution, plates and HETP.
- R_s = 2(4.8 − 4.2)/(0.30 + 0.32) = 1.2/0.62 = 1.94.
- N = 16(4.8/0.32)² = 16 × 225 = 3600.
- HETP = 30/3600 m = 8.3 mm.
Answer: R_s = 1.94 (baseline separated), N = 3600, HETP = 8.3 mm.
Example: A GC gives areas 1200, 800 and 500 for A, B, C with response factors 1.0, 1.2 and 0.9. Find the composition.
- Corrected areas: 1200, 960, 450; total 2610.
- A = 46.0 %, B = 36.8 %, C = 17.2 %.
Answer: A 46 %, B 37 %, C 17 % (mass %, if the factors are mass-based).
Spectroscopy, electrochemical and gas analysers
UV–visible spectrophotometry and colorimetry rest on the Beer–Lambert law A = log(I₀/I) = εbc: absorbance is proportional to concentration c and path length b. A spectrophotometer selects a wavelength with a monochromator (prism or grating) and measures transmitted light with a photodetector; a colorimeter uses filters, often after adding a reagent that develops colour (as for phosphate or iron in water). Concentrations are read from a calibration line made with standards.
Ion meters measure ion activity with ion-selective electrodes: the pH glass electrode develops a potential of about 59.16 mV per pH unit at 25 °C (Nernst equation E = E° − (2.303RT/F)pH) against a reference electrode; other membranes respond to Na⁺, F⁻, NO₃⁻ etc. Polarography applies a slowly increasing voltage to a dropping-mercury electrode; reducible species give current steps whose half-wave potential identifies them and whose limiting (diffusion) current is proportional to concentration.
Flue and fuel gas analysis: the Orsat apparatus absorbs CO₂ (KOH), O₂ (alkaline pyrogallol) and CO (cuprous chloride) from a measured dry sample; modern analysers use non-dispersive infrared (CO, CO₂), paramagnetic or zirconia cells for oxygen, and gas chromatography. Paramagnetic oxygen analysers exploit oxygen’s strong attraction to a magnetic field; zirconia probes produce a voltage from the oxygen-concentration difference across a hot ceramic electrolyte and sit directly in flue ducts to trim combustion air. Combustible-gas detectors use catalytic beads (pellistors) that heat up as flammable gas burns on them, or infrared absorption, reading in % of the lower explosive limit.
- Beer–Lambert: A = εbc = −log T.
- pH electrode: 59.16 mV/pH at 25 °C.
- Polarography: limiting current ∝ concentration.
- O₂: paramagnetic, zirconia; CO/CO₂: NDIR; Orsat for dry flue gas.
- Combustible gas: catalytic bead, % LEL.
Example: A sample transmits 40 % of the light in a 1 cm cell. Standards give A = 0.0500 per mg/L. Find the concentration.
- A = −log(0.40) = 0.398.
- c = 0.398/0.0500 = 7.96 mg/L.
Answer: About 8.0 mg/L.
Example: A pH electrode reads 177.5 mV more negative than in a pH 4 buffer at 25 °C. What is the sample pH?
- Slope = 59.16 mV per pH unit.
- ΔpH = 177.5/59.16 = 3.0.
- pH = 4 + 3 = 7.
Answer: pH 7.
GC (carrier gas, column, oven; TCD, FID, ECD) and HPLC separate mixtures; retention time identifies, peak area quantifies; N, HETP and resolution measure performance. Mass spectrometry identifies by m/z and fragments (GC–MS). UV–Vis and colorimetry use Beer–Lambert with calibration standards. Ion-selective electrodes and pH (Nernst slope 59.16 mV/pH). Polarography: diffusion current ∝ concentration. Gas analysis: Orsat, NDIR, paramagnetic and zirconia O₂, catalytic and IR combustible-gas detectors.
- FID does not respond to water, CO₂ or N₂.
- Keep absorbance roughly in 0.1–1 for best accuracy.
- Zirconia O₂ probes trim combustion air.
- Gas detectors alarm at 10–25 % LEL.
Key formulas
- R_s = 2(t₂ − t₁)/(w₁ + w₂)
- N = 16(t_R/w)²
- A = εbc = −log T
- E = E° − 0.05916 pH (25 °C)
Practice questions
The most suitable GC detector for hydrocarbons is
- TCD
- FID
- ECD
- UV
Show the answer
Answer: B. FID Flame ionisation.
Beer–Lambert law relates absorbance to
- temperature
- concentration and path length
- pressure
- flow
Show the answer
Answer: B. concentration and path length A = εbc.
Resolution of 1.5 between two peaks means
- complete overlap
- baseline separation
- no peaks
- one peak
Show the answer
Answer: B. baseline separation Standard criterion.
A zirconia probe measures
- CO₂
- oxygen
- methane
- moisture
Show the answer
Answer: B. oxygen Solid electrolyte.
Polarography uses a
- glass electrode
- dropping-mercury electrode
- thermocouple
- FID
Show the answer
Answer: B. dropping-mercury electrode Heyrovský.
Theory questions
Describe a gas chromatograph with a neat sketch and explain how a mixture is analysed quantitatively. (6 marks)
Show the model answer
Parts: carrier-gas cylinder with pressure/flow control; heated injection port (syringe or process sampling valve); column (packed or capillary with liquid stationary phase) in a programmable oven; detector (TCD/FID); data system. Components partition between the gas and stationary phase differently, travel at different speeds and emerge as separate peaks. Qualitative: retention time compared with standards. Quantitative: peak area proportional to amount; calibrate with standards to get response factors, then mass % = (A_iRF_i)/Σ(A_jRF_j) × 100 (normalisation) or use an internal standard. Column performance by plate number and resolution.
Explain the principle and working of a UV–Visible spectrophotometer. (5 marks)
Show the model answer
Light from a source (deuterium lamp for UV, tungsten for visible) passes through a monochromator (grating and slits) that selects a narrow wavelength band, then through the sample cuvette and onto a detector (photodiode or photomultiplier). A double-beam instrument splits the light through sample and reference cells to cancel drift. The instrument computes absorbance A = log(I₀/I). By Beer–Lambert, A = εbc, so a calibration line of A versus concentration from standards at the wavelength of maximum absorbance gives the unknown concentration. Deviations occur at high concentration, with stray light or chemical changes.
Level measurement
Topics: Direct measurement of liquid level: dipstick, hook type, sight glass, float type, displacer type; indirect measurement: hydrostatic head, bubbler (purge), capacitance, ultrasonic and radiation types.
Direct and indirect level measurement
Direct methods read the liquid surface itself. A dipstick or dip tape is lowered into the tank and the wetted length read; a hook gauge measures small level changes in open channels and weirs precisely. A sight (gauge) glass is a transparent tube or a reflex/transparent flat glass connected to the vessel at top and bottom; by communicating vessels the liquid stands at the same level. A float follows the surface and moves a pointer via a tape and counterweight, or operates a switch; a displacer is a heavy cylinder that does not float but loses weight by buoyancy as the level rises, measured by a torque tube or spring, giving a continuous signal suitable for interface measurement.
Indirect methods infer level from another quantity. Hydrostatic-head instruments measure pressure at the bottom: h = Δp/(ρg). In closed vessels a differential-pressure transmitter compares bottom pressure with the vapour space; if the reference (low-side) leg is filled with liquid (wet leg) the zero is suppressed or elevated accordingly. A bubbler (purge) system blows a small, steady flow of air or nitrogen through a dip tube; the back pressure needed equals the liquid head above the tube end, so the instrument never touches the liquid (good for slurries and corrosive liquids). Capacitance probes form a capacitor with the vessel wall: as the level rises the dielectric changes from vapour to liquid and the capacitance rises linearly. Ultrasonic (and radar) sensors on the tank top time an echo from the surface: distance = vt/2, level = height − distance. Radiation (gamma) gauges place a source on one side and a detector on the other: absorption by the liquid reduces the count rate, giving a completely non-contact measurement for very hot, toxic or high-pressure vessels.
- Direct: dipstick, hook, sight glass, float, displacer.
- Hydrostatic: h = Δp/ρg; wet-leg zero suppression/elevation.
- Bubbler: back pressure = head above tube end.
- Capacitance: C rises linearly with level.
- Ultrasonic: h = H − vt/2; radiation: non-contact through walls.
Example: A DP transmitter on an open water tank reads 25 kPa. Find the level.
- h = Δp/(ρg) = 25 000/(1000 × 9.81).
Answer: 2.55 m above the lower tap.
Example: An ultrasonic sensor 4 m above the tank bottom hears its echo after 10 ms (v = 343 m/s). Find the level.
- Distance to surface = 343 × 0.010/2 = 1.715 m.
- Level = 4 − 1.715 = 2.285 m.
Answer: 2.29 m.
Example: A capacitance probe (3 m) in a tank has D/d = 20; the liquid has ε_r = 2.2. The reading is 100 pF. Find the level.
- Capacitance per metre in vapour = 2πε₀/ln 20 = 5.563 × 10⁻¹¹/2.996 = 18.57 pF/m.
- C = 18.57[2.2h + (3 − h)] = 18.57(3 + 1.2h).
- 100/18.57 = 5.385 = 3 + 1.2h ⇒ h = 1.99 m.
Answer: About 1.99 m.
Flow and level measurement
Head-type flow meters (orifice, venturi) give Q ∝ √Δp, so a fourfold rise in differential pressure means twice the flow. Magnetic meters suit conductive liquids; rotameters are simple variable-area meters. Level: sight glass, float, differential pressure (Δp = ρgh), ultrasonic and radar.
- Q ∝ √Δp
- DP level: h = Δp/ρg
- Magnetic meter needs a conductive liquid
Example: An orifice’s differential pressure rises from 4 to 16 kPa. By what factor has the flow changed?
- Q ∝ √Δp, so the ratio is √(16/4).
- = 2: the flow has doubled.
Answer: Doubled
Example: A water tank is 2 m deep. What does a DP level transmitter at the bottom read?
- Δp = ρgh = 1000 × 9.81 × 2.
- ≈ 19.6 kPa.
Answer: ≈ 19.6 kPa
Direct level: dipstick, hook gauge, sight glass, float (tape, switch), displacer (buoyancy, torque tube, interfaces). Indirect: hydrostatic head with DP transmitters (wet/dry legs, zero suppression and elevation), bubbler/purge, capacitance probes, ultrasonic and radar echo time, gamma radiation absorption for difficult service.
- Hydrostatic methods need the density.
- Bubblers suit slurries and corrosive liquids.
- Foam and vapour disturb ultrasonic echoes; radar is better.
- Radiation gauges need licensing and shielding.
Key formulas
- h = Δp/(ρg)
- C = 2πε₀[ε_r h + (L − h)]/ln(D/d)
- h = H − vt/2
- Buoyancy on displacer = ρgA h
Practice questions
A displacer level instrument works on
- Bernoulli’s principle
- Archimedes’ principle
- Pascal’s law
- Ohm’s law
Show the answer
Answer: B. Archimedes’ principle Buoyancy.
A bubbler level system measures
- echo time
- back pressure of purge gas
- capacitance
- radiation count
Show the answer
Answer: B. back pressure of purge gas Equals liquid head.
For a hot, toxic, high-pressure vessel with no openings, the best choice is
- sight glass
- dipstick
- gamma radiation gauge
- float
Show the answer
Answer: C. gamma radiation gauge Non-contact through walls.
A wet reference leg on a DP level transmitter requires
- no correction
- zero suppression/elevation
- a float
- a thermocouple
Show the answer
Answer: B. zero suppression/elevation Constant head on low side.
Ultrasonic level measurement uses
- the speed of sound and echo time
- buoyancy
- thermal conductivity
- ionisation
Show the answer
Answer: A. the speed of sound and echo time h = H − vt/2.
Theory questions
Classify level measuring instruments and describe the hydrostatic and bubbler methods. (6 marks)
Show the model answer
Direct: dipstick/tape, hook gauge, sight glass, float-and-tape, float switches, displacer. Indirect: hydrostatic (pressure gauge, DP transmitter), bubbler/purge, capacitance, conductivity probes, ultrasonic, radar, radiation, weighing (load cells). Hydrostatic: bottom pressure p = ρgh plus vapour pressure; a DP transmitter connects its high side to the bottom tap and low side to the vapour space (dry leg) or a filled reference leg (wet leg, needing zero elevation); h = Δp/ρg, so density must be known and constant. Bubbler: a dip tube reaches near the bottom; air/N₂ is supplied through a restriction at a small constant flow; when bubbles escape, the tube pressure equals the liquid head above its end, read by a gauge or transmitter. No contact of the instrument with the liquid; suitable for corrosive liquids and slurries; needs a gas supply.
Explain capacitance and ultrasonic level measurement. (5 marks)
Show the model answer
Capacitance: an insulated probe and the vessel wall (or a concentric tube) form a capacitor; C = 2πε₀[ε_r h + (L − h)]/ln(D/d) increases linearly with the height of liquid h because the liquid’s permittivity exceeds the vapour’s. An RF bridge measures C. Conducting liquids need an insulated probe; coating and changing ε_r cause errors. Ultrasonic: a transducer on the tank top emits pulses (20–200 kHz) that reflect from the liquid surface; the transit time t gives distance vt/2 and level H − vt/2. Non-contact; the speed of sound is temperature-compensated; foam, vapour, dust and turbulence weaken echoes (radar is used then).
Surface area measuring instruments
Topics: BET method, scanning electron microscopy, dynamic vapour sorption, transmission electron microscopy.
BET adsorption and dynamic vapour sorption
Catalysts, adsorbents, cements and powders are often characterised by their specific surface area (m²/g), most of which lies inside pores. The BET method measures the amount of an inert gas, usually nitrogen at its boiling point 77 K, adsorbed on a degassed sample at a series of relative pressures x = P/P₀. Brunauer, Emmett and Teller extended Langmuir’s monolayer theory to multilayers and derived x/[V(1 − x)] = 1/(V_mC) + [(C − 1)/(V_mC)]x, which is a straight line over about 0.05 ≤ x ≤ 0.30. From the slope and intercept, the monolayer volume V_m = 1/(slope + intercept) and the constant C = slope/intercept + 1 (related to the heat of adsorption). The surface area is S = (V_m/22 414) N_A σ, with σ = 0.162 nm² for N₂. The full isotherm (adsorption and desorption with hysteresis) also gives pore volume and pore-size distribution (BJH method).
Dynamic vapour sorption (DVS) measures the mass of a sample on a sensitive microbalance while a carrier gas of controlled humidity (or organic vapour concentration) flows over it. Stepping the relative humidity up and down gives sorption–desorption isotherms and uptake kinetics, used for pharmaceuticals, foods, polymers and building materials (moisture uptake, hydrate formation, stability).
- BET line valid for P/P₀ ≈ 0.05–0.30.
- V_m = 1/(slope + intercept); C = slope/intercept + 1.
- S = (V_m/22 414) × N_A × 0.162 nm² ≈ 4.35 V_m m²/g.
- DVS: mass vs humidity on a microbalance.
Example: A BET plot gives slope 0.0450 g/cm³ and intercept 0.0004 g/cm³ (V in cm³ STP/g). Find V_m, C and the surface area.
- V_m = 1/(0.0450 + 0.0004) = 22.03 cm³/g.
- C = 0.0450/0.0004 + 1 = 113.5.
- S = 22.03/22 414 × 6.022 × 10²³ × 0.162 × 10⁻¹⁸ = 95.9 m²/g.
Answer: V_m ≈ 22.0 cm³/g, C ≈ 114, about 96 m²/g.
Example: In a DVS run a 20.00 mg sample weighs 21.30 mg at 80 % RH. Find the moisture uptake.
- Uptake = (21.30 − 20.00)/20.00 × 100.
Answer: 6.5 % by mass.
Electron microscopy: SEM and TEM
Light microscopes cannot resolve features much smaller than half the wavelength of light (about 0.2 µm). Electrons accelerated through a voltage V have a de Broglie wavelength λ ≈ 1.226/√V nm (non-relativistic; slightly less at high voltage), only picometres, so electron microscopes reach nanometre or even atomic resolution. Both types need a vacuum, an electron gun (tungsten, LaB₆ or field-emission), electromagnetic lenses and usually conductive or thin samples.
In the scanning electron microscope (SEM) a fine beam (typically 1–30 kV) is scanned across the surface; secondary electrons give topography with large depth of field, backscattered electrons give composition contrast, and characteristic X-rays analysed by energy-dispersive spectroscopy (EDS/EDX) give elemental composition. Resolution is about 1–10 nm, and particle size, shape and surface texture are seen directly. In the transmission electron microscope (TEM) a high-energy beam (typically 100–300 kV) passes through a very thin specimen (< 100 nm); the transmitted electrons form an image or diffraction pattern, revealing internal structure, crystal lattices, pores, metal particles on catalyst supports and defects, with resolution down to about 0.1 nm. Image analysis of many particles gives size distributions, from which a surface area can be estimated and compared with BET.
- λ ≈ 1.226/√V nm.
- SEM: surface topography, EDS composition, 1–10 nm.
- TEM: thin samples, internal structure, ~0.1 nm.
- Both need vacuum; non-conductors are coated (gold, carbon).
Example: Estimate the electron wavelength in an SEM at 10 kV and compare it with green light (550 nm).
- λ ≈ 1.226/√10 000 = 0.01226 nm.
- Ratio 550/0.01226 ≈ 45 000.
Answer: About 0.012 nm, some 45 000 times shorter than green light (lens aberrations, not wavelength, limit SEM resolution).
BET: N₂ adsorption at 77 K; linear BET plot for P/P₀ ≈ 0.05–0.30 gives V_m and C; surface area = V_m N_A σ/22 414 (σ = 0.162 nm²); isotherms also give pore size (BJH). DVS: gravimetric water/vapour sorption isotherms and kinetics. SEM: scanned beam, secondary and backscattered electrons, EDS, surface morphology at nm resolution. TEM: transmitted electrons through thin samples, internal and atomic structure.
- Degas samples before BET to clean the surface.
- C indicates adsorbate–surface interaction strength.
- SEM has large depth of field.
- TEM samples must be electron-transparent.
Key formulas
- x/[V(1 − x)] = 1/(V_mC) + [(C − 1)/(V_mC)]x
- S = V_m N_A σ/22 414
- λ ≈ 1.226/√V nm
- uptake % = (m − m₀)/m₀ × 100
Practice questions
BET surface area is usually measured with
- oxygen at 300 K
- nitrogen at 77 K
- helium at 4 K
- water at 25 °C
Show the answer
Answer: B. nitrogen at 77 K Liquid nitrogen temperature.
The BET equation is linear over P/P₀ of about
- 0–0.01
- 0.05–0.30
- 0.5–0.9
- 0.9–1
Show the answer
Answer: B. 0.05–0.30 Multilayer region.
The cross-sectional area of an adsorbed N₂ molecule is taken as
- 0.016 nm²
- 0.162 nm²
- 1.62 nm²
- 16.2 nm²
Show the answer
Answer: B. 0.162 nm² Standard value.
Elemental composition in an SEM is found by
- EDS (X-rays)
- secondary electrons only
- BET
- DVS
Show the answer
Answer: A. EDS (X-rays) Characteristic X-rays.
TEM requires samples that are
- thick
- very thin
- liquid
- magnetic
Show the answer
Answer: B. very thin Electron transparent.
Theory questions
Explain the BET method for surface area determination. (6 marks)
Show the model answer
The degassed sample is cooled in liquid nitrogen (77 K) and known doses of N₂ are admitted; the volume adsorbed V (cm³ STP/g) is measured at relative pressures P/P₀ from about 0.05 to 0.3 (and higher for a full isotherm). BET theory (multilayer adsorption, Langmuir assumptions for each layer, heat of adsorption of higher layers = heat of liquefaction) gives P/[V(P₀ − P)] = 1/(V_mC) + [(C − 1)/(V_mC)](P/P₀). Plotting the left side against P/P₀ gives a straight line; V_m = 1/(slope + intercept), C = slope/intercept + 1. Surface area S = V_m N_A σ/22 414 m²/g with σ(N₂) = 0.162 nm². Used for catalysts, adsorbents, cements, powders; pore-size distribution by BJH from the desorption branch.
Compare SEM and TEM. (5 marks)
Show the model answer
SEM: focused beam (1–30 kV) scanned over the surface; image from secondary (topography) and backscattered (atomic-number contrast) electrons; bulk samples, coated if non-conducting; large depth of field, 3-D appearance; resolution ≈ 1–10 nm; EDS for composition. TEM: broad high-energy beam (100–300 kV) transmitted through a thin (< 100 nm) specimen; image formed by lenses from transmitted electrons; shows internal structure, lattice fringes, defects, nanoparticles; diffraction patterns give crystal structure; resolution ≈ 0.1 nm; demanding sample preparation (thin sections, microtomy, ion milling).
Instrumentation symbols, PFDs and P&IDs
Topics: Instrumentation symbols; introduction to process flow diagrams and piping and instrumentation diagrams.
ISA symbols and tag numbers
Instruments on drawings follow ISA 5.1 (and IS/ISO equivalents). Each instrument is a circle (bubble) containing a tag: letters describing its function and a loop number. The first letter is the measured or initiating variable: A analysis, F flow, L level, P pressure, T temperature, S speed, W weight, H hand. Succeeding letters give the function: I indicate, R record, C control, T transmit, E primary element, A alarm (with H/L for high/low), S switch, V valve, Y relay or computing function, Z final element. So FT-101 is a flow transmitter in loop 101, TIC-205 a temperature indicating controller, LAH a high-level alarm, PSV a pressure safety valve, FE an orifice plate.
The bubble’s style shows where the instrument is: a plain circle is field-mounted; a circle with a solid horizontal line is in the main control room (panel or console accessible to the operator); a dashed line means behind the panel. A circle in a square is a shared display/control (DCS function), a hexagon a computer function, a diamond in a square a PLC function. Signal lines: solid thick for process piping; thin solid with double cross-hatches for pneumatic signals (older drawings use slashes); dashed for electrical signals; dash–dot or with small circles for software/data links; capillary and hydraulic lines have their own patterns. Control valves are drawn as two triangles (bow-tie) with an actuator symbol (diaphragm dome, piston, motor, solenoid) and a fail action (FO, FC, FL).
- First letter: measured variable (F, L, P, T, A…).
- Next letters: function (I, R, C, T, E, A, S, V).
- Circle: field; line inside: control room; square around: DCS.
- Dashed line: electric signal; hatched: pneumatic.
Example: Decode the tags FIC-101, LT-202, TAH-305, PSV-410 and FE-101.
- FIC-101: flow indicating controller, loop 101.
- LT-202: level transmitter, loop 202.
- TAH-305: temperature alarm high, loop 305.
- PSV-410: pressure safety valve.
- FE-101: flow element (e.g. orifice plate) feeding loop 101.
Answer: Read the first letter as the variable and the rest as functions.
From PFD to P&ID
A process flow diagram (PFD) shows the major equipment with standard symbols and tag numbers, the main process and utility streams with numbers and a stream table (flow, composition, temperature, pressure), heat duties and the principal control loops. It is used for process design, material and energy balances and equipment sizing.
A piping and instrumentation diagram (P&ID) is the detailed engineering document: every pipe with line number, size, material specification and insulation; every manual and control valve, reducer, strainer and blind; every instrument with its tag, bubble type and signal lines; control loops, interlocks and alarms; safety devices (relief valves, rupture discs, flame arrestors); vents, drains, sample points and utility connections. It is the basis for construction, hazard and operability (HAZOP) studies, commissioning, operation and maintenance. A typical control loop on a P&ID reads: a sensor/element (FE) and transmitter (FT) in the field, a controller (FIC) in the DCS comparing the measurement with the set point, and a control valve (FV) with its actuator and fail position.
- PFD: equipment, streams, stream table, main controls.
- P&ID: lines, valves, instruments, interlocks, safety devices.
- P&ID is the basis for HAZOP.
- Loop: element → transmitter → controller → valve.
Example: Sketch in words the P&ID for controlling the level of a reflux drum by manipulating the reflux flow.
- LT-101 (field circle) on the drum, connected by dashed electric line to LIC-101 (circle with line, in DCS).
- LIC-101 output (dashed line) goes to LV-101, a control valve with diaphragm actuator on the reflux line, fail-open or fail-closed by safety analysis.
- Add LAH-101 and LAL-101 alarms and a local level gauge LG-101.
Answer: LT-101 → LIC-101 → LV-101 on the reflux line, with high and low alarms and a gauge glass.
ISA 5.1 symbols: tag letters (first = measured variable, succeeding = functions) and loop numbers; bubble styles for field, control-room, behind-panel, DCS, computer and PLC; line types for process, pneumatic, electric and software signals; control-valve and actuator symbols with fail positions. PFD: equipment, numbered streams, stream table, duties, main controls. P&ID: all lines, valves, instruments, loops, interlocks and safety devices; basis for HAZOP and construction.
- Same loop number links all instruments of a loop.
- Fail position chosen for safety (e.g. fuel valve fail-closed).
- PSV set pressure ≤ design pressure.
- P&IDs are revised and controlled documents.
Key formulas
- Tag = variable letter + function letters + loop number
- Loop: FE → FT → FIC → FV
Practice questions
In the tag TIC, the letter T stands for
- transmitter
- temperature
- time
- total
Show the answer
Answer: B. temperature First letter = variable.
A circle with a horizontal line through it means the instrument is
- field mounted
- in the main control room
- behind the panel
- a PLC
Show the answer
Answer: B. in the main control room ISA 5.1.
A dashed signal line represents
- process pipe
- electrical signal
- pneumatic signal
- capillary
Show the answer
Answer: B. electrical signal Electric.
Which document is the basis for a HAZOP study?
- BFD
- PFD
- P&ID
- plot plan
Show the answer
Answer: C. P&ID Shows every valve and instrument.
PSV stands for
- pressure safety valve
- process solenoid valve
- pump suction valve
- pressure switch
Show the answer
Answer: A. pressure safety valve Relief device.
Theory questions
Explain the ISA instrument identification system with examples. (5 marks)
Show the model answer
Each instrument has a tag of letters and a loop number. First letter: measured/initiating variable (A analysis, F flow, L level, P pressure, T temperature, S speed, W weight, Q quantity, H hand). Optional modifier (D differential, F ratio, Q totalise). Succeeding letters: readout/passive function (I indicate, R record, G glass/gauge, A alarm, E element, L light) and output function (C control, T transmit, S switch, V valve, Y relay/compute, Z final element), with H/L modifiers. Examples: PDT differential pressure transmitter; FRC flow recording controller; LSHH level switch high-high; TE temperature element; AIT analyser indicating transmitter. Bubble styles indicate location (field, control room, DCS, PLC).
Distinguish between a PFD and a P&ID and list the contents of each. (5 marks)
Show the model answer
PFD: conceptual/process design document. Contents: major equipment with symbols and tags, main process lines with flow direction and stream numbers, stream table (flow, composition, T, P, phase), utility streams, heat duties, key operating conditions and main control loops; no pipe sizes or minor valves. P&ID: detailed engineering document. Contents: all equipment with design data, every pipe with line number, size, spec and insulation, all valves (manual, control, check, relief), reducers, strainers, instruments with ISA tags and signal lines, control loops, interlocks, alarms, safety devices, vents, drains, sample points, utility connections, tie-ins and notes. Used for detailed design, HAZOP, construction, operation and maintenance.
Notes, examples and questions were written for this studio; the prescribed books are listed in the studio for reference.
Frequently asked questions
What is the Process Instrumentation Studio?
Measure it right: sensors, signals and symbols. Static and dynamic characteristics, first- and second-order response, transducers, thermocouples and RTDs, pressure and vacuum gauges, chromatography and spectroscopy, level, BET and electron microscopy, ISA symbols and P&IDs from the ASTU CHE241405 syllabus.
What can I do in the Process Instrumentation Studio?
First- and second-order instrument response; Thermocouples, RTDs, pyrometers and pressure gauges; Chromatography, spectroscopy and level measurement; BET surface area, SEM/TEM and P&ID symbols.
Is the Process Instrumentation Studio free?
Yes. It is free and runs in your browser with no sign-up.